Horizon Med Sci 2017, 23(2): 105-109 Back to browse issues page
Comparison the Precision of Ear and Finger Pulse Oximetry with Finger Sensor in Mechanically Ventilated ICU Patients
T. Baloochi Beydokhti1, A. Mohamadpour1, M. kianmehr2, MJ shaban *3
1- Medical-Surgical Nursing Department, Nursing & Midwifery Faculty, Gonabad University of Medical Sciences, Gonabad, Iran
2- Medical Physics Department, Medicine Faculty, Gonabad University of Medical Sciences, Gonabad, Iran
3- Student Research Committee, Gonabad University of Medical Sciences, Gonabad, Iran , shabanmj1@mums.ac.ir
Abstract:   (349 Views)

Aims: Nowadays, the pulse oximetry is widely used to monitor the oxygen saturation constantly. In the clinical care, it is observed that the ear pulse oximetry with the finger sensor is sometimes utilized in the ICUs. The aim of the study was to determine the accuracy of the ear pulse oximetry and the fingertip with the finger sensor in the ICU patients under the mechanical ventilation. 

Materials & Methods: In the analytic comparative study, 60 hospitalized patients under the mechanical ventilation in the ICU of Moddares Hospital in Kashmar Township were studied in 2015. The subjects were selected via available convenience sampling method. One finger pulse oximetry sensor having been attached to the upper part of the ear and the other to the fingertip, the arterial blood sampling was done simultaneously. Data was analyzed by SPSS 22 software using repeated ANOVA and Spearman correlation tests. 

Findings: Mean oxygen saturation percentage and the heartbeat of ear, fingertip, and arterial blood pulse oximetry were significantly different (p<0.001). In addition, the oxygen saturation percentage of ear, fingertip, and arterial blood pulse oximetry was significantly correlated by the partial pressure of oxygen (PaO2; p=0.001). Nevertheless, the mentioned parameters were not significantly correlated by the peripheral edema and especially, by the hand (p>0.05).   

Conclusion: The utilization of ear pulse oximetry with the finger sensor on the fingertip cannot lead to accurate results of the oxygen saturation in the hospitalized patients in the ICU under the mechanical ventilation. However, in case of any reduction in the arterial blood oxygen pressure, the accuracy of the finger pulse oximetry starts to drop out.

Keywords: Oximetry, Arterial Oxygen Saturation, Intensive Care Units
Full-Text [PDF 447 kb]   (295 Downloads)    
Type of Study: Applicable | Subject: Anesthesiology
Received: 2016/10/17 | Accepted: 2017/01/25 | Published: 2017/03/11
Send email to the article author



DOI: 10.18869/acadpub.hms.23.2.105


XML   Persian Abstract   Print


Download citation:
BibTeX | RIS | EndNote | Medlars | ProCite | Reference Manager | RefWorks
Send citation to:

Baloochi Beydokhti T, Mohamadpour A, kianmehr M, shaban M. Comparison the Precision of Ear and Finger Pulse Oximetry with Finger Sensor in Mechanically Ventilated ICU Patients. Horizon Med Sci . 2017; 23 (2) :105-109
URL: http://hms.gmu.ac.ir/article-1-2619-en.html
Volume 23, Number 2 (Spring 2017) Back to browse issues page
مجله علمی پژوهشی افق دانش The Horizon of Medical Sciences
Persian site map - English site map - Created in 0.05 seconds with 783 queries by yektaweb 3503